Safety/Regulatory Information
1
-16
0993-5255
Safety practices for flame atomization
Safety interlocks
PerkinElmer provides a number of safety interlocks on the burner system to
monitor gas pressure and check for the proper setup of the burner head, nebulizer,
and drain system. In addition a flame sensor checks that the flame is burning.
Do not attempt to defeat these interlocks; you may compromise your own, or
someone else’s, safety.
Safe use of burner gases
The seepage of fuel gas or fuel gas mixture from the burner system, the drain
system, the gas control system, or the gas connections constitutes a serious fire
hazard.
Make sure that there are no breaks or leaks in any of these systems and that
all the seals are correctly installed and in good condition.
Regularly check for leaks at joints and seals using an approved leak test
solution.
When you perform any maintenance or installation procedures, follow the
instructions in the guide exactly.
Do not attempt to service the gas control system yourself.
A PerkinElmer service engineer, or similarly authorized and trained person,
must perform the work.
When you shut down the instrument, for example at the end of the working
day, shut all the gas lines at the cylinder or regulator valves. Bleed the lines
between the regulator and instrument to atmosphere before switching off the
ventilation system.
Содержание AAnalyst 700
Страница 1: ...ATOMIC SPECTROMETRY User s Guide AAnalyst 700 Atomic Absorption Spectrometer Return to Document Menu...
Страница 2: ......
Страница 3: ...ATOMIC SPECTROMETRY AAnalyst 700 Atomic Absorption Spectrometer User s Guide...
Страница 8: ...Contents C 4 0993 5255...
Страница 10: ...1 2 0993 5255...
Страница 44: ...Safety Regulatory Information 1 36 0993 5255...
Страница 46: ...Spectrometer Installation 2 2 0993 5255...
Страница 68: ...Spectrometer Installation 2 24 0993 5255...
Страница 70: ...Preparing the System 3 2 0993 5255...
Страница 84: ...System Maintenance 4 2 0993 5255...
Страница 98: ...Parts and Supplies 5 2 0993 5255...
Страница 102: ...Laboratory Requirements 6 2 0993 5255...
Страница 120: ...Spectrometer System Description 7 2 0993 5255...
Страница 136: ...Translations of Warnings 8 2 0993 5255...
Страница 146: ...Translations of Warnings 8 12 0993 5255...
Страница 185: ...0993 5255 In 1 Index...
Страница 186: ...Index In 2 0993 5255...
Страница 192: ......
Страница 193: ......