73
archive data). To do this, insert the SD/SDHC card in the transducer
slot (contacts facing down) and make sure that the card has been
properly mounted (the top right corner of the display shows
a card icon
). The percent value of archive use, at which the data
will automatically be copied to the card or to the file internal file sy
-
stem memory, must be set. This value is placed in register 7614 or can
be changed using menu:
Archive
→
Param.SD
. For example if
“20.0” is entered in register 7614, data will be collected in the internal
transducer memory until the use of the internal memory reaches 20%,
then the automatic archive copying to the SD/SDHC card or the file
system internal memory process will begin. If the percent value of use
will be higher, e.g. 99%, then data will be written on the SD/SDHC card
less frequently, but the writing process will take longer. Writing data
to the card is indicated with a progress bar graph displayed at the lo-
wer line of the LCD display. Do not remove the SD/SDHC card from
the transducer slot if writing to the card is in progress, because this
could lead to data corruption or device reset. Writing can be stopped
and the card can be removed once it is unmounted (section 5.3.2).
It is also possible, to force archive copy to the SD/SDHC card or
file system internal memory at any time by pressing the combina
-
tion of
keys. If the transducer is in the variant
with the Ethernet interface, the archive data can be downloaded from
the file system memory via the FTP protocol using any FTP client
software.
Note:
If the transducer is connected to the FTP client, then copying
the archive data from the internal memory to file system memory is
blocked! In order to acquire current data from the archive, disconnect
the FTP session and force archive copy (e.g. press
keys). After copying is finished connect again transducer with FTP client
software.
5.8.3. Archiving configuration
Registers 4064-4069 (table 37) and transducer menu
in
Settings
→
Archive
group (table 7) are used for configu
-
ring archiving parameters. The archiving can be constant or conditional.