33
6.0 Manual Mode
Manual mode allows you to operate the VitaPump manually. You can select pump speeds from 0-
100%, and set the direction, CW or CCW. In Manual mode, gravimetric metering (with a balance) is
not possible. This is useful when priming the system is desired, or when the balance has failed and
you need to run in manual mode. When metering in Manual mode, be certain to verify the speed
required to obtain the correct addition rate.
From the Mode Select screen, scroll to Manual and press “Select”. You will see the following
screens:
Mode Select: MANUAL
Up
Down
Select
-MANUAL-
100.0%
SW=0000
CW
Use the
RATE
key to increase or decrease the motor speed.
Use the
Double Arrow key
to set or change the pump direction.
Use the
RUN
and
STOP
keys to start and stop the pump.
The
SWITCH
key provides access to 4 TTL outputs not available in other modes.
The
TIME
and * keys do not function on the VitaPump.
The
External RUN/STOP
still functions normally in Manual mode.