4.10 Setup: Scale
4.11 Mettler - Toledo Balance Parameters:
4.111 Mettler Viper PM, PB Models
To enter the Technical Setup mode, press and hold the “Print” key until the word
CODE
appears on the display. Then press “Zero”, “Tare”, “Zero”, “Tare”, “Print”, and
the
word
SCALE
will appear. You are at the top of the menu. Use the “Zero” and
“Tare” buttons to scroll thru choices on a level, and the “Print” button to select.
Menu Level 1
Level 2
Level 3 Selected Menu
Item SCALE
1.1 Serial Number (snr)
1.2 Metrology(metrolo)
1.3 Build
Unit
Resolution
g
0.1g
1.4 Linearity (linear)
1.5 Geo Value (geo)
1.6 Calibration (cal)
1.8 Units
Displayed
g (grams)
1.9 Tare
1.10 Zero
4.112 Mettler ICS425 series models
To enter the Setup mode, press and hold the “Transfer” key until the word CODE
appears on the display. Then press “Zero”, “Zero”, “Zero”, “Transfer”, and an
interactive menu screen is presented. Use the displayed arrows and the related
buttons to navigate the menu and confirm the following settings are enabled.
Level 1
Level 2
Level 3
Selected Menu Item
SCALE
Disp Unit & res Disp Unit 1
g
Disp Unit 2
g
Disp resolution OFF (sets to default value)
Zero AZM
OFF
Filter Vibration
Medium
Process
Dosing
Stability
Standard
Comm COM1
Mode
Dialog
Param
RS232
Baud
9600
Parity
8
none
H.Shake
Off
When exiting the Menu a prompt to save occurs, press OK to save the changes.
NOTE:
The parameters listed are the only custom settings needed for communication with Parker
systems. All others should remain at factory defaults. Consult your scale manual for help navigating
through the scale menu. If stability is an issue, adjust the Vibration and Stability settings as needed.
In Setup Mode, select “METLER”, see SETUP: SCALE: MANUF. By making this selection, the
system will implement the correct parameters for communicating with this scale. You will also need
P/N 080-072ICS Cable.
Содержание VitaPump
Страница 1: ...VitaPump System Installation Operation Maintenance Instructions ...
Страница 19: ...19 ...
Страница 21: ...21 2 0 Vitamin Metering Edit Alarm ...
Страница 23: ...23 3 0 Metering Exec ...
Страница 24: ...24 4 0 Setup Edit ...
Страница 34: ...34 Appendix A Tandem Pump Head Installation and Maintenance ...
Страница 37: ...37 TANDEM Dual Channel Peristaltic Pump Parts Identification ...
Страница 39: ...39 Appendix B FMI Q1 Pump Head Installation and Maintenance ...
Страница 40: ...40 ...
Страница 41: ...41 ...
Страница 42: ...42 ...
Страница 43: ...43 ...
Страница 44: ...44 ...