Note 1: At 80 Hz and 800 MHz, the higher frequency range applies.
Note 2: This guidance may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic environment
– Guidance
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2,5 GHz
10 V/m
Recommended separation
distance:
d = 0.35 * SQRT (P)
d = 0.35 * SQRT (P) 80 MHz to
800 MHz
d = 0.70* SQRT (P) 800 MHz to
2.5 GHz
Where „P“ is the maximum out-
put power rating of the transmit-
ter in watts (W) according to the
transmitter manufacturer and
„d“ is the recommended sepa-
ration distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site
survey
a)
, should be less than
the compliance level in each
frequency range
b)
.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele-
phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which the
eFlow
®
rapid
is used exceeds the applicable RF compliance level above, the eFlow
®
rapid
should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as reorient-
ing or relocating the eFlow
®
rapid
.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Table 2 and 4 of DIN EN 60601-1-2:2007 (IEC 60601-1-2:2007)
Содержание eflowrapid 178G1005
Страница 1: ...Gebrauchsanweisung Instructions for Use nebuliser system...
Страница 6: ...2 eFlow rapid 03 12...
Страница 38: ...34 eFlow rapid 03 12...
Страница 70: ...66 eFlow rapid 03 12...
Страница 102: ...98 eFlow rapid 03 12...
Страница 134: ...130 eFlow rapid 03 12...
Страница 166: ...162 eFlow rapid 03 12...
Страница 198: ...194 eFlow rapid 03 12...
Страница 230: ...226 eFlow rapid 03 12...
Страница 262: ...258 eFlow rapid 03 12...
Страница 294: ...290 eFlow rapid 03 12...
Страница 329: ......