Electromagnetic immunity
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic environment
– Guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
transient /burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input /
output lines
±2 kV for power
supply lines
±1 kV for input /
output lines
Mains power quality should be
similar to that of a typical com-
mercial or hospital environ-
ment.
Surge
IEC 61000-4-5
±1 kV common
mode
±2 kV differential
mode
±1 kV common
mode
±2 kV differential
mode
Mains power quality should be
similar to that of a typical com-
mercial or hospital environ-
ment.
Voltage dips,
short interrup-
tions and volt-
age variations
on power sup-
ply input lines
IEC 61000-4-11
<5% U
T
a)
(>95% dip of U
T
)
for 1/2 cycle
40% U
T
(60% dip of U
T
)
for 5 cycles
70% U
T
(30% dip of U
T
)
for 25 cycles
<5% U
T
(>95% dip of U
T
)
for 5 s
a) U
T
is the mains voltage prior to application of the test level.
<5% U
T
(>95% dip of U
T
)
for 1/2 cycle
40% U
T
(60% dip of U
T
)
for 5 cycles
70% U
T
(30% dip of U
T
)
for 25 cycles
<5% U
T
(>95% dip of U
T
)
for 5 s
Mains power quality should be
similar to that of a typical com-
mercial or hospital environ-
ment. When the user of the
eFlow
®
rapid
continued function
also calls in the event of disrup-
tion of supply, it is recom-
mended the eFlow
®
rapid
from
an uninterruptible power supply
or a battery.
Power fre-
quency (50 Hz/
60 Hz) mag-
netic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels char-
acteristic of a typical location in
a typical commercial or hospital
environment.
Conducted RF
IEC 61000-4-6
3 V
eff
150 kHz to
80 MHz
10 V
eff
150 kHz to
80 MHz
Portable and mobile RF com-
munication equipment should
be used no closer to any part of
the eFlow
®
rapid
, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Содержание eflowrapid 178G1005
Страница 1: ...Gebrauchsanweisung Instructions for Use nebuliser system...
Страница 6: ...2 eFlow rapid 03 12...
Страница 38: ...34 eFlow rapid 03 12...
Страница 70: ...66 eFlow rapid 03 12...
Страница 102: ...98 eFlow rapid 03 12...
Страница 134: ...130 eFlow rapid 03 12...
Страница 166: ...162 eFlow rapid 03 12...
Страница 198: ...194 eFlow rapid 03 12...
Страница 230: ...226 eFlow rapid 03 12...
Страница 262: ...258 eFlow rapid 03 12...
Страница 294: ...290 eFlow rapid 03 12...
Страница 329: ......