Temperature calibration procedure
:
chamber sensor
)
Cal key is vialed within setting range, touch
Cal key.
Move to Menu display, touch Service #2
key
.
Move to Service display, touch Temp.
Calibration key
Move to input value display.
Input measured value by thermometer
Touch mark area.
Touch TOP key, so return to Menu display.
Input value area return to zero.
NOTE:
:
Ambient Temp. sensor and condenser A&B sensor can calibrate in the same procedure.
Keep touching MENU KEY (5 sec)
Move to Service Code input display,
so input “335232”
-
43.3
-
43.3
-
43.3
-20-