Operation
How to operate Service # 1 and Service # 2 on the service menu is described below. For other operations on the LCD, refer
to the instruction manual.
CAUTION) Firmware update, compressor, FAN, and cap heater settings affect basic performance.
Be careful when operating.
Term
Function
Runtime
Display the accumulate time of Main battery, Backup battery FAN A and FAN B, Reset
them.
Version UP
Update the firmware of main board and touch panel.
Temp.
Calibration
Chamber adjustment method.
System Configuration #1
Setting Capi. Heater and its timer, compressors, Fan A/B .
Status
Monitoring of chamber temp. , ambient temp, comp. evaporator and fan.
System Configuration #2
Turning On/O
ff
of AIP heater, and Backup. Setting overload ala
r
m delay..
Reset Data
Initialization of data, and flash memory. Setting model code.
Error Log
Output of error log data.
Setup Data
Input and output the data.
Other Setting
Turning On/O
ff
Screen Capture, Demo Mode, Self-diagnosis Function and FAN Alarm.