SECTION 10 SERIAL REMOTE INTERFACE
120
10.3 INSTALLATION
Installation of the RS-232 interface is generally performed at the factory. This involves removing the
IEEE-488 interface PCB assembly and rear panel connector/cable and installing the RS-232 interface
assembly and rear panel connector/cable.
INTERFACE PCB DIP Switch S1-5 must be set to ON to indicate that the RS-232 interface has been
installed.
10.4 SETUP
After installing RS-232 interface into the UPC (or after erasing all program memory or after system
RESET), the RS-232 communication parameters must be set (BAUD RATE, PARITY and EOS
Terminator) as follows:
1. Press
the
fn
key, select
GENERAL SETUP
then
UPC STATUS
.
2. Press
the
EDIT
key and enter the desired
BAUD RATE
numerically, selected from the list
above. Then press
ENTER
.
3. Select
the
PARITY
by pressing the
+/-
key until the desired selection (NONE, ODD or EVEN) is
shown. Then press
ENTER
.
4. Select
EOS
(End Of String) Terminator by pressing the
+/-
key until the desired selection
(CR/LF, LF or CR) is shown. Then press
ENTER
.
5. Press
STORE
.
NOTE:
Incorrect communication parameters may result in the UPC displaying a:
PARITY ERROR,
FRAMING ERROR, and/or
OVERRUN ERROR.
Sending a BREAK signal will result in a BREAK RCVD message and will terminate data transmission in
progress from the UPC.
10.5 OPERATION
Operation of the UPC via RS-232 REMOTE interface consists of utilizing the same commands available
for the IEEE-488 interface except as noted below. (see SECTION 5)
NOTE:
Normally the UPC is in the LISTENER mode. When the UPC is Queried for information, the UPC
automatically enters the Talker mode until the response data is sent or another command is sent from
the Host.
Содержание UPC-12
Страница 2: ......
Страница 12: ...SECTION 1 GENERAL viii THIS PAGE INTENTIONALLY BLANK ...
Страница 17: ...SECTION 1 GENERAL 5 PACIFIC Figure 1 3a SCU UPC 32 Figure 1 3b UPC 32 Figure 1 3c UPC 12 ...
Страница 46: ...SECTION 4 OPERATION 34 PA CIFIC FIGURE 4 1 POWER SOURCE FRONT PANEL ...
Страница 77: ...SECTION 4 OPERATION 65 90 0 100 0 85 95 100 90 FIGURE 4 6 2 1 EDITED WAVEFORM ...
Страница 139: ...NOTES 127 NOTES ...
Страница 140: ...MODIFICATIONS 128 MODIFICATIONS MODIFICATIONS This section is reserved for any modifications that may be made ...