SECTION 8
CALIBRATION
113
8.3
EXTERNALLY REFERENCED CALIBRATION (cont’d)
The UPC has 2 separate current metering circuits, one for FORM 1 (single phase), and another for FORM 2 and
FORM 3. In order to fully calibrate all current meters for all power forms, this procedure must be performed
twice.
UPC-12 a. Operate the power source in FORM 1 to calibrate
I1
(single phase).
b. Operate the power source in FORM 2 to calibrate
Ia
and
Ib
(single/split phase).
UPC-32 a. Operate the power source in FORM 1 to calibrate
I1
(single phase).
b. Operate the power source in FORM 3 to calibrate
Ia, Ib
and
Ic
(three phase).
Only the phases existent need to have values entered. If a phase parameter is not used or is not available as in
the case of a 1 or 2 phase power source, the phase C parameters should be set to 0. Entering 0 for a parameter
causes the correction factor (kFactor) for that parameter to be set to the system default value, 1.0000 .
NOTE:
Only
I1
or
Ia, Ib
and
Ic
can be entered depending upon the power form currently in use.
NOTE:
Voltages are to be entered as AC RMS,
LINE TO NEUTRAL only for voltages
.
Currents are to be entered as AC RMS.
Each time a non-zero value is entered, the UPC measures the output signal for that parameter and compares it
against the entered external meter calibration value. A new calibration factor (k factor) is then calculated and
used to correct the meter display for that value.
If a value for a parameter is not keyed in, the existing calibration factor will remain in memory and will not be re-
calculated. The correction factors are stored in Battery-backed memory (RAM) until the memory is lost or
cleared.
Содержание UPC-12
Страница 2: ......
Страница 12: ...SECTION 1 GENERAL viii THIS PAGE INTENTIONALLY BLANK ...
Страница 17: ...SECTION 1 GENERAL 5 PACIFIC Figure 1 3a SCU UPC 32 Figure 1 3b UPC 32 Figure 1 3c UPC 12 ...
Страница 46: ...SECTION 4 OPERATION 34 PA CIFIC FIGURE 4 1 POWER SOURCE FRONT PANEL ...
Страница 77: ...SECTION 4 OPERATION 65 90 0 100 0 85 95 100 90 FIGURE 4 6 2 1 EDITED WAVEFORM ...
Страница 139: ...NOTES 127 NOTES ...
Страница 140: ...MODIFICATIONS 128 MODIFICATIONS MODIFICATIONS This section is reserved for any modifications that may be made ...