SECTION 4
OPERATION
49
4.5.1.1 PROGRAM EXAMPLE
The following is an example of how to create and execute a 3 phase, 108/187 volt, 400 Hz program.
From any metering display press the
PROGRAM
key.
PROGRAM: #1 FORM=3 COUPLING=DIRECT
FREQ=60 Va=120 Vb=120 Vc=120
WFa=1 WFb=1 WFc=1
#SEGS=0 Ilim=50 PHb=120º PHc=240º
Press
4,
ENTER, EDIT,
to edit program #4.
PROGRAM: #4 FORM=3 COUPLING=DIRECT
FREQ=60 Va=120 Vb=120 Vc=120
WFa=1 WFb=1 WFc=1
#SEGS=0 Ilim=50 PHb=120º PHc=240º
Press
3,
ENTER,
to select 3 phase form.
PROGRAM: #4 FORM=3 COUPLING=DIRECT
FREQ=60 Va=120 Vb=120 Vc=120
WFa=1 WFb=1 WFc=1
#SEGS=0 Ilim=50 PHb=120º PHc=240º
Press
ENTER,
to select DIRECT coupling. If XFMR is shown press
+/-,
then
ENTER
.
PROGRAM: #4 FORM=3 COUPLING=DIRECT
FREQ=60 Va=120 Vb=120 Vc=120
WFa=1 WFb=1 WFc=1
#SEGS=0 Ilim=50 PHb=120º PHc=240º
Press
400, ENTER,
to select 400 Hz..
PROGRAM: #4 FORM=3 COUPLING=DIRECT
FREQ=400 Va=120 Vb=120 Vc=120
WFa=1 WFb=1 WFc=1
#SEGS=0 Ilim=50 PHb=120º PHc=240º