OxyFlo Pro User Manual
Revision 1.1 | May 2015
42
If the problem persists please contact technical support.
The monitor is receiving power (the magenta power-on indicator light
on top of the monitor is on) but is not booting to the usual ‘default’
screen
Turn off the universal AC adapter, wait 2-3 minutes and try again. If the
problem persists there is likely to be a fault, in which case please contact
technical support.
The monitor boots into the usual ‘default’ screen but does not respond
to the connection of a probe
Try power-cycling the monitor by turning off the AC adapter, waiting 2-3
minutes and turning the AC adapter (i.e. the monitor) back on. Try another
probe and/or try switching channels. If the problem persists there is likely to
be a fault, in which case please contact technical support.
The monitor boots into the usual ‘default’ screen but the display does
not respond to any touch inputs
Try power-cycling the monitor by turning off the AC adapter, waiting 2-3
minutes and turning the AC adapter (i.e. the monitor) back on. Try
connecting a probe in one or more channels. If the problem persists there is
likely to be a fault, in which case please contact technical support.
There is a continuous sound upon power-on
The monitor has failed the power-on self-test. If the problem repeats itself
after multiple power-on attempts contact technical support.
There is an audible ‘beep’ every 5 seconds and a warning triangle has
appeared in the bottom right of the display which displays a
‘Temperature Out of Range’ message when selected
If the internal temperature of the OxyFlo Pro monitor rises above the
maximum permitted internal operating temperature threshold then the
message ‘Temp out of Range’ will appear on the display. If this occurs
check that the fan at the rear of the OxyFlo Pro is not obstructed and/or
move the instrument to a cooler location. Analogue outputs will continue but
the trend values will not appear on the display. Data generated during this
CAUTION
Connecting the NX PSU AC adapter to the
DC POWER IN
inlet at the rear of the OxyFlo Pro, having
already
powered it
up via its On/Off switch MAY trigger the surge protection
circuit within the OxyFlo Pro. If the OxyFlo Pro does not start
up, turn off the AC adapter, wait 2 - 3 minutes, then turn the
AC adapter on once more.