OxyFlo Pro User Manual
Revision 1.1 | May 2015
23
5. Power up the monitor via the On/Off switch on the NX-PSU AC
adapter. The AC adapter will automatically detect the local mains
voltage (110/120V or 220/240V supported). The red, ring-shaped
power-on indicator light on top of the monitor will illuminate and the
monitor will generate an audible ‘beep’. While the monitor runs its
internal boot sequence and start-up checks (approx. 15 seconds) the
display will initially remain blank. There will be two further audible
‘beeps’ approx. 5 seconds apart, at which point the Oxford Optronix
logo will be briefly displayed before the screen launches into its
‘default’ view (below).
6. The monitor is now ready for probe connection and use (see section
3.4).
CAUTION
Connecting the NX-PSU universal AC adapter to the
DC
POWER IN
inlet at the rear of the OxyFlo Pro, having
already
powered it via its On/Off switch MAY trigger the surge
protection circuit within the OxyFlo Pro. If the OxyFlo Pro
does not start up, turn off the AC adapter, wait 2 - 3 minutes,
then turn the AC adapter on once more.