A visual reminder is displayed that the probe needs to be in the
calibration solution, which will be accompanied by an audible double
‘beep’ every 2 seconds.
7. Touch the CAL button a second time to trigger the calibration. A 5s
continuous ‘beep’ is emitted, followed by the following (animated)
graphic which confirms that the calibration is progressing.
IMPORTANT: This 20 - 30 second period is where any vibration
or movement will cause the calibration procedure to fail
.
If the calibration is successful the monitor will emit an audible double
‘beep’ and briefly display the following on-screen confirmation:
The display will then automatically revert to the default view (the
warning triangle now absent) and will display a reading from the
calibration solution, which should be in the region of 1,000 BPU (+/-
50 BPU).
Re-calibration of Probes
Probes can be re-calibrated at any time as above, via the ‘CAL’ function
available on the probe status dialogue view.
Failed Calibration
A failed calibration will be indicated by a long continuous ‘beep’ followed by
a series of ‘beeps’ corresponding to the error number identified. The probe