OxyFlo Pro User Manual
Revision 1.1 | May 2015
24
3.4
Connecting / Disconnecting Probes
The OxyFlo Pro automatically recognises previously calibrated probes and
applies the necessary probe calibration coefficients on connection. This
essentially alleviates the requirement for repeated re-calibration.
Figure 6: Connecting and disconnecting a probe
Align the probe connector such that the arrow is aligned with the black dot on
the instrument connector surround and simply push. The connector will lock
into position, the probe ID will be read and calibration coefficients loaded
automatically by the monitor (provided the probe has been either factory-
calibrated or calibrated by the user – see instructions below).
Disconnect by twisting the front portion of the probe connector anti-clockwise,
in the direction of the ‘Release’ arrow printed on the connector.
NOTE: OxyFlo Pro monitors shipped
prior to May 2015
will only function if
pairs
of connectors are occupied. An optional ‘interlock adapter’ may be
used to occupy the unused channel where the user wishes to use an odd
number of probes (one or three).
CAUTION
Attempting to disconnect the probe by pulling the cable
sleeving instead of the probe connector may cause
irreparable damage to the probe.