OxyFlo Pro User Manual
Revision 1.1 | May 2015
31
Keep the active probe surface or probe tip away from the edges or
the bottom of the bottle. The probe should be supported in such a
way that it does not swing or move whilst in the solution.
4. Connect the probe that you wish to calibrate into the requisite
channel of the monitor (ensuring that the paired channel is occupied
either by another probe or by an LDF Interlock Adapter).
NOTE: If the probe has not previously been calibrated in this channel
the toolbar will display a warning triangle
and
a red cross.
5. Select the probe from the toolbar to call up the probe status
dialogue. If the probe has not previously been calibrated in this
channel the status dialogue will display no date and a red cross
against the calibration data indicator, as follows:
6. Touch the ‘CAL’ button to initiate the calibration routine.