OxyFlo Pro User Manual
Revision 1.1 | May 2015
11
1.11
Summary of Cautions for Probe Calibration
CAUTION
It is essential that the calibration procedure is performed on a
stable and vibration-free surface. Any movement or vibration
during the calibration procedure - however slight - is likely to
result in a failed calibration.
CAUTION
The motility standard has a limited life. The expiry date is
indicated on the label. The solution must not be used beyond
this date, as it may produce misleading values due to the
gradual aggregation of the latex spheres.
CAUTION
Do not use the motility standard in ambient temperatures
below 15ºC or above 25ºC.
CAUTION
Store the motility standard within the temperature range 3 –
25ºC. DO NOT FREEZE the solution.
CAUTION
Do not dilute the motility standard.
CAUTION
It is important that all probes to be used with any one OxyFlo
Pro monitor have
different probe identification numbers
in
order to prevent possible probe calibration errors. Please
contact technical support for advice.