6.3.5
Stopping a sweep
1
Tap
Control
on the Home page. The Control Loop Configuration page is displayed.
2
Select the required temperature-sensor in the
Sensor
field.
3
Tap the Set
(K)
button until it displays
Manual
. The
Set (K)
field value is held at its
present value.
6.4
Using the plot function
6.4.1
Configuring a plot
1
Tap
Plot
on the Home page. The Plot Configuration page is displayed.
2
Tap
Device
and select the first device to be used.
3
Tap
Signal
and select the first signal to be plotted.
4
Tap
Save
to plot the first signal.
5
To add traces to the plot, double-tap anywhere on the plot area and repeat steps
6
If required, tap once in the area below the plot to add a scale on the X axis (see
).
Figure 6-1
Signal Monitor Plot page with X and Y axis scales