GETTING STARTED
Original Instructions
5
GETTING STARTED
Ensure that you have read and understood the information in
before installing
or operating this equipment.
5.1
Checking the electrical supply
The iTC automatically configures itself to match the supplied electrical power supply, if
the voltage and frequency are within the specified ranges (see
intervention is required.
5.2
Mounting the equipment
The equipment is housed inside a 19 inch, 2U rack-mountable enclosure. The enclosure
can either be placed on a bench or a table, or fitted within a suitable rack enclosure.
Ensure there is sufficient access at the rear for the electrical connections.
If the iTC is mounted in a rack, ensure that the rear of the rack is open. Site the rack at
least 30 cm from a wall or other obstruction, to ensure adequate airflow around the unit.
Caution
CHECK THE SUPPLY VOLTAGE
Check that the voltage supplied to the equipment is within the specified
range. Operating the equipment with an incorrect voltage may cause
permanent damage.
CONTRÔLER LA TENSION D'ALIMENTATION
Contrôler que la tension fournie à l'équipement est comprise dans les
limites spécifiées. Exploiter l'équipement avec une tension incorrecte
peut l'endommager de manière permanente.
Caution
EQUIPMENT VENTILATION
If the Mercury unit is to be rack mounted, to ensure an adequate airflow
around the unit it is important that the back of the rack is open and the
rack is not placed less than 30 cm from a back wall.
Lorsque le Mercury est monté en rack, s'assurer que le rack est ouvert à
l'arrière et se trouve au minimum à 30 cm du mur afin de garantir une
circulation d'air suffisante.