Handbook
November 2011
/
59-UMC0071
/
Issue
01
Mercury
iTC
Page 65
FUNCTIONAL DESCRIPTION
Original Instructions
4.3.2.1
Interconnecting heater and temperature-sensor boards
A heater board is always associated with a temperature sensor board. The output of a
heater board is hard-wired to the 9-way D-connector on an associated temperature
controller board. These connections are listed in
If one or more heater boards are fitted, temperature sensor boards must be fitted in the
corresponding slots listed in
.
A control-loop must be configured before a heater board can be operated. Any
temperature sensor can be configured to control any heater, as long as the hardware is
connected to match this configuration.
For example:
a) If a single heater board is fitted in slot 2, a temperature sensor board must be
fitted in slot 7. The control loop must be configured so that the temperature
sensor in slot 7 controls the heater in slot 2.
Table 4-5
Locations of temperature sensor and heater boards
Heater Board Output
Temperature Sensor
Board
Slot 1
is connected to
Slot 6
Slot 2
is connected to
Slot 7
Slot 3
is connected to
Slot 8