Handbook
November 2011
/
59-UMC0071
/
Issue
01
Mercury
iTC
Page 138
LEVEL-METER DAUGHTER BOARD
Original Instructions
16.3
Configuring the level-meter board
If you are using the level-meter with liquid helium, see
If you are using the level-meter with liquid nitrogen, see
16.3.1
Configuring a liquid-helium sensor
1
Tap an unconfigured panel on the Home page. The Channel Display Configuration
page is displayed.
2
Tap on the
Devices
field and select a level meter input.
3
Tap on the
Signals
field and select
Resistance
.
4
Tap
Assign
to save the changes.
5
Tap
Home
.
6
With the helium sensor at room temperature, record the sensor resistance value
that is displayed on the Home page.
7
Fully immerse the sensor part of the probe in liquid helium.
8
Record the sensor resistance value that is displayed on the Home page. This value
should be close to zero.
9
Tap the configured level-meter panel on the Home page. The Level Meter Sensor
Details page is displayed.
10
Select the
Helium
tab.
Figure 16-3 Level Meter Sensor Details page - Helium tab
11
Enter the resistance measured in step
Resistance at 0% (
Ω
)
field.
12
Enter the resistance measured in step
Resistance at 100% (
Ω
)
field.
13
Enter a value between 0 and 160 mA in the
Measurement Pulse Amplitude (mA)
field.
14
Enter a value in seconds in the
Measurement Pulse Duration (s)
field.
15
Tap the
Fast/Slow
button until it displays
Slow
.