GETTING STARTED
Original Instructions
5.3
Connecting electrical power
Connect the equipment to an electrical supply, using an appropriate mains cable.
In the EU, the mains supply cable must be rated for the maximum current rating of the
equipment. The cable must meet the requirements of IEC227 or IEC245. Mains cables
certified or approved by any recognised nation test house are regarded as meeting this
requirement.
In the USA, the mains cable must be a listed mains cable set approved to the standard
UL817 for cable sets and power supply cables. The mains plug must be a separable
plug without a locking device. If the mains cable is to be a disconnecting device for the
equipment, the cable must be shorter than 3 m. The cable set must be suitable for
purpose and must have a current rating at least 125% that of the equipment rating.
In Canada, the cable set must be certified by CSA.
Green/yellow covered conductors must only be used for connection to protective ground
terminals.
5.3.1
Connecting the heaters and sensors
1
Connect all sensors as described in
.
2
Connect all heaters as described in
5.3.2
Connecting the data interfaces
If it is required to connect the iTC to a computer, refer to the relevant section below.
5.3.2.1
Connecting the ethernet interface (if required)
1
Connect the ethernet interface to the computer, using a standard ethernet cable.
5.3.2.2
Connecting the RS232 serial interface (if required)
1
Connect the RS232 interface to the computer, using a suitable cable (See
5.3.2.3
Connecting ISOBUS (if required)
1
Connect the iTC to one of the Slave connectors on the ISOBUS cable.
2
Connect the computer to the Master connector on the ISOBUS cable.