S-178-10(E)
(3)
電池の保存寿命は出荷後約 10 年です。
通常は外部から電源供給して電池をほとんど消耗しない場合でも、保存寿命により約 10 年で電池の交換が
必要となります。(従って、前頁計数部の寿命とあわせて計数部一式の交換となります。)保存寿命は特に
下記のような環境では更に短くなる場合がありますので、早めの交換を推奨します。
◎高温雰囲気
◎高温流体計測
◎寒冷地
Storage life of the battery is 10 years approx after shipment.
Normally, even in an application where the register is powered by an external source and the battery
appears virtually free from any current drain, the storage life requires replacement in ten years approx.
(Taking the abovementioned life expectancy of the register, replacing the entire register assembly will
thus be required.) The battery may possibly have even a shorter storage life under any of the following
environments, requiring early replacement:
◎
High temperature environment
◎
High temperature fluid measurement
◎
Cold region
症 状
1. 通液しない。
原 因
1. 流量計のフランジ入口と出口の
保護カバーを外さないで配管した。
. オーバル回転子にスケールがかみ込
み回転しなくなり、計量液が流れない
と考えられる場合。
. ポンプ圧力またはヘッド圧が不足。
対 策
1. 流量計を外し、カバーを除去。
. 計数部を分離し、本体部の分解、
洗浄の実施。
. 配管系全体の圧力損失を考慮し適
正なポンプを選定。
〈配管要領 1 頁参照〉
8.故 障 対 策