96
Prosthesis battery
Battery type
Li-Ion
Charging cycles (charging and discharging
cycles) after which at least 80% of the original
battery capacity remains available
500
Charge level after 1 hour charging time
30 %
Charge level after 2 hours charging time
50 %
Charge level after 4 hours charging time
80 %
Charge level after 8 hours charging time
Fully charged
Product behaviour during the charging process The product is non-functional
Operating time of the prosthesis with new, fully
charged battery at room temperature
Approx. 5 days with average use
Power supply
Reference number
757L16*
Storage and transport in original packaging
-40 °C/-40 °F to +70 °C/+158 °F
Storage and transport without packaging
-40 °C/-40 °F to +70 °C/+158 °F
10 % to 93 % relative humidity, non-condens
ing
Operation
0 °C/+32 °F to +40 °C/+104 °F
max. 90 % relative humidity, non-condensing
Input voltage
100 V~ to 240 V~
Mains frequency
50 Hz to 60 Hz
Output voltage
12 V
Battery charger
Reference number
4E60*
Storage and transport in original packaging
-25 °C to 70 °C/-13 °F to 158 °F
Storage and transport without packaging
-25 °C to 70 °C/-13 °F to 158 °F
Max. 93% relative humidity, non-condensing
Operation
5 °C to 40 °C/41 °F to 104 °F
Max. 93% relative humidity, non-condensing
Protection rating
IP40
Input voltage
12 V
Only when "DUAL" marking is not present (
see Page 60
)
Cockpit app
Article number
Cockpit 4X441-Andr=V*
Supported operating system
Android 4.0.3 or higher
Website for download
http://www.ottobock.com/cockpitapp
Data transfer
Wireless technology
Bluetooth 2.0
Range
approx. 10 m / 32.8 ft
Frequency range
2402 MHz to 2480 MHz
Modulation
GFSK
Содержание 3B5-X2
Страница 2: ...2...
Страница 56: ...56...
Страница 61: ...61 Using the 4X350 X3 remote control is no longer possible on products with the marking DUAL...