99
17.2 Operating states/error signals
The prosthesis indicates operating states and error messages through beeps and vibration sig
nals.
17.2.1 Signals for operating states
Battery charger connected/disconnected
Beep signal
Vibration signal
Event
—
3x long
Charging mode started (3 sec. after connecting
the battery charger)
1x short
1x short
Self-test completed successfully, product is
operational
Mode switching
INFORMATION
When mute mode (silent mode) is activated, no beep and vibration signals are generated.
INFORMATION
If the
Volume
parameter is set to '0' in the Cockpit app, there are no beep signals (see
Page 83).
Beep signal
Vibration
signal
Additional action per
formed
Event
1x short
1x short
Mode switching using the
Cockpit app
Mode switching is performed
using the Cockpit app.
1x long
1x long
Bouncing on the forefoot fol
lowed by unloading the
prosthetic leg
Bouncing pattern recognised.
1x short
1x short
Weight taken off prosthetic
leg and leg kept still for 1
second
Switching to basic mode
(mode 1) carried out.
2x short
2x short
Weight taken off prosthetic
leg and leg kept still for 1
second
Switching to MyMode 1 (mode 2)
carried out.
3x short
3x short
Weight taken off prosthetic
leg and leg kept still for 1
second
Switching to MyMode 2 (mode 3)
carried out.
17.2.2 Warnings/error signals
Error during use
Beep signal
Vibration signal
Event
Required action
–
1 x long at intervals of
approx. 5 seconds (if
mute mode (silent
mode) is activated,
this signal is not gen
erated)
Overheated hydraulic
unit
Reduce activity.
Содержание 3B5-X2
Страница 2: ...2...
Страница 56: ...56...
Страница 61: ...61 Using the 4X350 X3 remote control is no longer possible on products with the marking DUAL...