102
Status
LED
Charging status
indicator (5 LEDs)
Error
Resolution
The LED
ring is
flashing
red
The prosthesis is not
being charged
Defective battery
charger.
Reset the error by dis
connecting and con
necting the power sup
ply.
If the error persists, the
battery charger and
power supply must be
inspected by an
authorised Ottobock
Service Center.
17.2.3 Error messages while establishing a connection with the cockpit app
Error message
Cause
Correction
Component was con
nected to another
device. Establish con
nection?
The component was
connected to another
device
To disconnect the original connection, tap
the "
OK
" button.
If the original connection is not to be dis
connected, tap the "
Cancel
" button.
Mode change failed
An attempt was made to
switch to a different
MyMode while the com
ponent was in motion
(e. g. while walking)
For safety reasons, switching MyModes is
only permitted when components are at
rest, e. g. while standing or sitting.
A current connection to
the component was
interrupted
Check the following points:
•
Distance from the component to the
device
•
Charge level of the component's bat
tery
•
Bluetooth of the component switched
on? (Switching Bluetooth of the com
ponent on/off)
•
Hold the component with the sole of the
foot facing up to make the component
"visible" for 2 minutes.
•
If multiple components were stored,
was the correct component selected?
17.2.4 Status signals
Battery charger connected
LED on
power
supply
Status
LED on
battery
charger
Event
Power supply and battery charger operational. Battery charger not yet
connected to receiver.