USER MANUAL
OMX 103UNI
| 35
5.
MENU
LIGHT
MEA
SURING MODE
T/C
38
→
→
Parameter
Display User
Setting
Description
Temperature of cold
junction
C.J.1
-20
…
99
Setting of cold junction temperature
•
Range: -20…99°C
•
DEF
0
Example setting
Temperature of cold junction 12°C >
12
0
›
›
1
›
›
2
›
›
0
2
›
›
1
2
›
›
FO.A
0
Parameter
Display User
Description
Decimal point
FO.A
Setting of decimal point. It can be either fixed, or floating (most
accurate projection format)
000
000
Integer
00,0
00.o
One decimal place
0,00
0.oo
Two decimal places
Floating decimal
point
FL.P.
Floating decimal point
DEF
Example setting
With a floating decimal point >
FL.P.
FL.P.
›
›
SET.
* The following menu item is dynamic and it depends on options fitted
i
Menu items “
CO.1
” and “
C.J.1
” are not available for T/C type "B".
The methodology and procedure for setting the cold junction
is described in a separate chapter. page <?>
000
00.o
0.oo
FL.P.
→
→
FO.A
C.J.1