USER MANUAL
OMX 103UNI
| 61
6.
MENU
PROFI
6.2
Setting - CHANNELS
6.2.1a Setting fi lters in channel A
2s
› › CHA. ›
› C. A ›
› FI.A ›
› MD.A ›
MD.A
Parameter
Display User
Setting
Description
Selecting filters for
input 1
MD.A
Sometimes it may be desirable to appropriately
mathematically adjust the measured data before
displaying it. The following filters are designed
to do just that.
Not applied
OFF
Filter is disabled
DEF
Average
AVR.
2
…
100
Average calculated from measured values
•
arithmetic average calculated from a specified
number of samples taken from
C. A
•
Range: 2…100
Floating filter
FLO.
2
…
30
Floating filter
•
floating arithmetic average calculated from a
specified number of samples taken from
C. A
recalculated with every new incoming sample
•
Range: 2…30
Exponential filter
EXP..
2
…
100
Exponential filter
•
First order integration filter with time constant
C. A
measurement
•
Range: 2…100
Rounding
RND.
2
…
100
Rounding of the measured value
•
it is set by any K. number which determines
the display step
•
e.g:
C. A
= 2,5 > display 0, 2.5, 5,...
•
Range: 0…999
→
→
MD.A
M.FN.
CHA.
SER.
OUT.
C. A
C. B
INP.
+
M. M.
Input parametrs are set in this menu
C. A
Setting the parameters of
“Channel A”
C. B
Setting the parameters of
"Channel B”
M.FN.
Setting the parameters of
math functions
M.M.
Selecting the channel from
which Min/Max values will
be evaluated