USER MANUAL
OMX 103UNI
| 53
6.
MENU
PROFI
2s
› INP. ›
› › › IN.1 ›
› › › CO.1 ›
CO.1
Parameter
Display User Description
Cold junction
CO.1
Cold junction compensation
(CJC)
•
when T. 1 > T/C
On terminals
IN.1
Measurement without a referential thermocouple
•
measurement of cold junction on connector terminals
DEF
→
→
On terminals
with ref. T/C
IN.2
Measurement with a referential thermocouple
•
CJC is measured on device's connector terminals with a
referential thermocouple in antiserial connection
→
→
Without ref. T/C
EX.1
Measurement without a referential thermocouple
•
entire set up is operated in homogenous and constant
temperature
→
→
Compensation
box
EX.2
Measurement with a referential thermocouple
•
when Compensation box is used
→
→
CO.1
i
There is more inofrmation on CJC configuration in the chapter
dedicated to CJC on page 86
For T/C type "
B"
menu items
CO.1
and
C.J.1
are NOT accessible
6.1.3d Cold junction compensation for input 1
T/C
2s
› INP. ›
› › › IN.1. ›
› › › › C.J.1 ›
C.J.1
Parameter
Display User
Setting
Description
Cold junction com-
pensation
0
-20
…
99
Same CJC configuration applies to Channel
1 and Channel 2
•
Range: -20…99°C
•
DEF
0°C
→
→
C.J.1
6.1.3e Additional resistance for input 1
RTD
OHM
2s
› INP. ›
› › › IN.1 ›
› › › › R.A.1 ›
R.A.1
Parameter
Display User
Setting
Description
Additional resistance
OHM, RTD
0
0
…
999
when it is necessary to offset the beginning
of the range by certain value, e.g. while using
sensor in measuring head to compensate for
leads' resistance
•
Range: 0…999 Ω
•
DEF
0 Ω
→
→
R.A.1