66
SP600 Spectrophotometer 01/2008
1.1 Methods
Chlorine, differentiated determination
with Powder Pack
0.01 – 2 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with
10 ml of the water sample
,
close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add
one VARIO Chlorine FREE-DPD / F10 powder
pack
straight from the foil to the water sample.
6. Close the vial tightly with the cap and swirl several
times to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that
the marks are aligned.
8. Press
TEST
key.
9. Remove the vial from the sample chamber. Empty the
vial, rinse vial and cap several times and then fill the vial
with
10 ml of water sample
.
10. Add
one VARIO Chlorine TOTAL-DPD / F10 powder
pack
straight from the foil to the water sample.
11. Close the vial tightly with the cap and swirl several
times to mix the contents (approx. 20 seconds).
prepare Zero
press ZERO
Zero accepted
prepare T1
press TEST
1
2
3
4
1
1
0
Содержание SP600
Страница 1: ...Instruction Manual SP600 Spectrophotometer ...
Страница 2: ......
Страница 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Страница 37: ...35 SP600 Spectrophotometer 01 2008 ...
Страница 73: ...71 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Oxidizing agents interfere as they react like Chlorine ...
Страница 83: ...81 SP600 Spectrophotometer 01 2008 ...
Страница 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 203: ...201 SP600 Spectrophotometer 01 2008 ...
Страница 207: ...205 SP600 Spectrophotometer 01 2008 ...
Страница 231: ...229 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Only ortho Phosphate ions react 2 P PO4 P2 O5 ...
Страница 245: ...243 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will be given ...
Страница 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 251: ...249 SP600 Spectrophotometer 01 2008 ...
Страница 266: ...264 SP600 Spectrophotometer 01 2008 ...
Страница 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Страница 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...
Страница 333: ...331 SP600 Spectrophotometer 01 2008 ...