325
SP600 Spectrophotometer 01/2008
Display
Possible Causes
Elimination
???
Example 1
0.60 mg/l free Cl
??? comb Cl
0.59 mg/l total Cl
Example 2
Underrange
??? comb Cl
1.59 mg/l total Cl
Example 3
0.60 mg/l free Cl
??? comb Cl
Overrange
The calculation of a value
(e.g. combined Chlorine) is
not possible
Test procedure correct?
If not – repeat test
Example 1:
The readings for free and total
Chlorine are different, but
considering the tolerances of
each reading they are the same.
For this reason the combined
Chlorine is most likely zero.
Example 2:
The reading for free Chlorine
is under the detection limit.
The instrument is not able
to calculate the combined
Chlorine. In this case the
combined Chlorine is most likely
the same as the total Chlorine.
Example 3:
The reading for total Chlorine
is exceeding the range.
The instrument is not able
to calculate the combined
Chlorine. The test should be
repeated with a diluted sample.
Error absorbance
e.g.: T2>T1
calibration of Fluoride was
not correct
Repeat calibration
Printer “timeout”
printer switched off;
no connection
Connect printer
Check connections
Switch printer on
Содержание SP600
Страница 1: ...Instruction Manual SP600 Spectrophotometer ...
Страница 2: ......
Страница 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Страница 37: ...35 SP600 Spectrophotometer 01 2008 ...
Страница 73: ...71 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Oxidizing agents interfere as they react like Chlorine ...
Страница 83: ...81 SP600 Spectrophotometer 01 2008 ...
Страница 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 203: ...201 SP600 Spectrophotometer 01 2008 ...
Страница 207: ...205 SP600 Spectrophotometer 01 2008 ...
Страница 231: ...229 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Only ortho Phosphate ions react 2 P PO4 P2 O5 ...
Страница 245: ...243 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will be given ...
Страница 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 251: ...249 SP600 Spectrophotometer 01 2008 ...
Страница 266: ...264 SP600 Spectrophotometer 01 2008 ...
Страница 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Страница 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...
Страница 333: ...331 SP600 Spectrophotometer 01 2008 ...