300
SP600 Spectrophotometer 01/2008
Performance of measurement:
Press [MODE] [5] [4] keys.
Confirm with
[ ]
key.
The display shows:
Enter the required wavelength in the range from 330 to 900 nm,
e.g. [4][0][0]. Confirm with
[ ]
key.
Afterwards the following queries are displayed:
Enter the time delay in the range from 0 to 999 sec.,
e.g. [6] [0]. Confirm with
[ ]
key.
Enter the interval time in the range from 6 to 999 sec.,
e.g. [2] [0]. Confirm with
[ ]
key.
< Kinetic>
wavelength:
_ _ _ nm
Interval tine:
_ _ _ s
< Kinetic>
Time delay:
_ _ _ s
Mode
5
4
4 0 0
6 0
2 0
Kinetic
With method 54 Kinetic it is possible to describe the time dependency of a reaction (e.g. time
of colour development). The maximum number of measuring points is 199 with an interval
time of 6 to 999 seconds. Concentrations of unknown samples can be calculated by a known
factor. The factor calculation is placed right before the sample measurement. To evaluate the
factor the slope of a best fit straight line through the measured Absorbance values is used.
For future measurements the factor can be entered by an option. If the sample concentration
is not of interest the factor shall be entered with one.
Attention:
If many points in short time intervals are measured a heating of the vial is
possible!
Содержание SP600
Страница 1: ...Instruction Manual SP600 Spectrophotometer ...
Страница 2: ......
Страница 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Страница 37: ...35 SP600 Spectrophotometer 01 2008 ...
Страница 73: ...71 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Oxidizing agents interfere as they react like Chlorine ...
Страница 83: ...81 SP600 Spectrophotometer 01 2008 ...
Страница 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 203: ...201 SP600 Spectrophotometer 01 2008 ...
Страница 207: ...205 SP600 Spectrophotometer 01 2008 ...
Страница 231: ...229 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Only ortho Phosphate ions react 2 P PO4 P2 O5 ...
Страница 245: ...243 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will be given ...
Страница 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 251: ...249 SP600 Spectrophotometer 01 2008 ...
Страница 266: ...264 SP600 Spectrophotometer 01 2008 ...
Страница 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Страница 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...
Страница 333: ...331 SP600 Spectrophotometer 01 2008 ...