21
SP600 Spectrophotometer 01/2008
1.1 Methods
Notes:
1. Before using clean the vials and the measuring beaker with Hydrochloric acid
(approx. 20%). Rinse then thoroughly with deionized water.
2. To get accurate results the sample temperature must be between 20°C and 25°C.
3. A low test result may be given in the presence of Fluorides and Polyphosphates.
The effect of this is generally insignificant unless the water has fluoride added
artificially. In this case, the following table should be used:
Fluoride
[mg/l F]
Displayed value: Aluminium [mg/l Al]
0.05
0.10
0.15
0.20
0.25
0.30
0.2
0.05
0.11
0.16
0.21
0.27
0.32
0.4
0.06
0.11
0.17
0.23
0.28
0.34
0.6
0.06
0.12
0.18
0.24
0.30
0.37
0.8
0.06
0.13
0.20
0.26
0.32
0.40
1.0
0.07
0.13
0.21
0.28
0.36
0.45
1.5
0.09
0.20
0.29
0.37
0.48
---
Example: If the result of Aluminium determination is 0.15 mg/l Al and the Fluoride
concentration is known to be 0.4 mg/l F, the true concentration of Aluminium is
0.17 mg/l Al.
4. Al
Al
2
O
3
Содержание SP600
Страница 1: ...Instruction Manual SP600 Spectrophotometer ...
Страница 2: ......
Страница 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Страница 37: ...35 SP600 Spectrophotometer 01 2008 ...
Страница 73: ...71 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Oxidizing agents interfere as they react like Chlorine ...
Страница 83: ...81 SP600 Spectrophotometer 01 2008 ...
Страница 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 203: ...201 SP600 Spectrophotometer 01 2008 ...
Страница 207: ...205 SP600 Spectrophotometer 01 2008 ...
Страница 231: ...229 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Only ortho Phosphate ions react 2 P PO4 P2 O5 ...
Страница 245: ...243 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will be given ...
Страница 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 251: ...249 SP600 Spectrophotometer 01 2008 ...
Страница 266: ...264 SP600 Spectrophotometer 01 2008 ...
Страница 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Страница 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...
Страница 333: ...331 SP600 Spectrophotometer 01 2008 ...