SP600 Spectrophotometer 01/2008
1.1 Methods
Notes:
1. Extremely basic or acidic water samples should be adjusted with 0.5 mol/l (1 N) Sulfuric
acid solution or 1 mol/l (1 N) Sodium hydroxide solution to pH 7.
2. Interferences:
Interfering substance
Interference levels and treatments
Calcium
greater than 1000 mg/l CaCO
3
Iron
Interferes at all levels. Correct as follows:
a) determine the concentration of iron present in the
sample by performing a total Iron test
b) add the same iron concentration as determined
to the deionized water (step 1).
The interference will be blanked out successfully.
Magnesium
greater than 6000 mg/l CaCO
3
Nitrate
greater than 100 mg/l NO
3
-N
Nitrite
greater than 12 mg/l NO
2
-N
Phosphate
greater than 100 mg/l PO
4
-P
Sulfate
greater than 300 mg/l SO
4
Sulfide
intensifies the colour
Glycine, Hydrazine,
Colour, Turbidity
Less common interferences such as Hydrazine and
Glycine will cause intensified colours in the prepared
sample. Turbidity and colour will give erroneous high
values. Samples with severe interferences require
distillation.
3. N
NH
4
NH
3
Содержание SP600
Страница 1: ...Instruction Manual SP600 Spectrophotometer ...
Страница 2: ......
Страница 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Страница 37: ...35 SP600 Spectrophotometer 01 2008 ...
Страница 73: ...71 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Oxidizing agents interfere as they react like Chlorine ...
Страница 83: ...81 SP600 Spectrophotometer 01 2008 ...
Страница 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 203: ...201 SP600 Spectrophotometer 01 2008 ...
Страница 207: ...205 SP600 Spectrophotometer 01 2008 ...
Страница 231: ...229 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Only ortho Phosphate ions react 2 P PO4 P2 O5 ...
Страница 245: ...243 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will be given ...
Страница 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 251: ...249 SP600 Spectrophotometer 01 2008 ...
Страница 266: ...264 SP600 Spectrophotometer 01 2008 ...
Страница 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Страница 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...
Страница 333: ...331 SP600 Spectrophotometer 01 2008 ...