48
SP600 Spectrophotometer 01/2008
1.1 Methods
Chloride
with Reagent Test
5 – 60 mg/l Cl
–
Use two clean vials (24 mm Ø) and mark one as blank for
zeroing.
1. Fill a clean vial (24 mm Ø) with
10 ml of deionized
water
(this is the blank).
2. Fill the second clean vial (24 mm Ø) with
1 ml of the
water sample
and
9 ml of deionized water
(this is
the sample).
3. Fill each vial with the same size drops by holding the
bottle vertically and squeeze slowly:
3 drops Chloride-51
4. Close the vials with the caps tightly and invert the vials
several times to mix the contents.
5. Fill each vial with the same size drops by holding the
bottle vertically and squeeze slowly:
3 drops Chlorid-52
6. Close the vials with the caps tightly and invert the vials
several times to mix the contents.
7. Press
[ ]
key.
Wait for a
reaction period of 3 minutes
.
8. Place the vial (the blank) in the sample chamber making
sure that the marks are aligned.
9. Press
ZERO
key.
10. Remove the vial from the sample chamber.
11. Place the vial (the sample) in the sample chamber ma-
king sure that the marks are aligned.
12. Press
TEST
key.
The result is shown in the display in mg/l Chloride.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
3:00
start:
9
1
Содержание SP600
Страница 1: ...Instruction Manual SP600 Spectrophotometer ...
Страница 2: ......
Страница 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Страница 37: ...35 SP600 Spectrophotometer 01 2008 ...
Страница 73: ...71 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Oxidizing agents interfere as they react like Chlorine ...
Страница 83: ...81 SP600 Spectrophotometer 01 2008 ...
Страница 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 203: ...201 SP600 Spectrophotometer 01 2008 ...
Страница 207: ...205 SP600 Spectrophotometer 01 2008 ...
Страница 231: ...229 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Only ortho Phosphate ions react 2 P PO4 P2 O5 ...
Страница 245: ...243 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will be given ...
Страница 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 251: ...249 SP600 Spectrophotometer 01 2008 ...
Страница 266: ...264 SP600 Spectrophotometer 01 2008 ...
Страница 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Страница 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...
Страница 333: ...331 SP600 Spectrophotometer 01 2008 ...