30
SP600 Spectrophotometer 01/2008
1.1 Methods
Ammonium HR
with Tube Test
1 – 50 mg/l N
1. Open one white capped reaction vial and add
0.1 ml
deionized water
(this is the blank).
2. Open another white capped reaction vial and add
0.1 ml of water sample
(this is the sample).
3. Add
one Vario AMMONIA Salicylate F5 powder
pack
straight from the foil into each vial.
4. Add
one Vario AMMONIA Cyanurate F5 powder
pack
straight from the foil into each vial.
5. Close the vials tightly and swirl several times to dissolve
the powder.
6. Press
[ ]
key.
Wait for a
reaction period of 20 minutes
.
After reaction period is finished proceed as follows:
7. Place the vial (the blank) in the sample chamber making
sure that the marks are aligned.
8. Press
ZERO
key.
9. Remove the vial from the sample chamber.
10. Place the vial (the sample) in the sample chamber
making sure that the marks are aligned.
11. Press
TEST
key.
The result is shown in the display in mg/l Ammonium.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
20:00
start:
1
2
3
4
Ø 16 mm
6
6
Содержание SP600
Страница 1: ...Instruction Manual SP600 Spectrophotometer ...
Страница 2: ......
Страница 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Страница 37: ...35 SP600 Spectrophotometer 01 2008 ...
Страница 73: ...71 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Oxidizing agents interfere as they react like Chlorine ...
Страница 83: ...81 SP600 Spectrophotometer 01 2008 ...
Страница 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 203: ...201 SP600 Spectrophotometer 01 2008 ...
Страница 207: ...205 SP600 Spectrophotometer 01 2008 ...
Страница 231: ...229 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Only ortho Phosphate ions react 2 P PO4 P2 O5 ...
Страница 245: ...243 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will be given ...
Страница 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 251: ...249 SP600 Spectrophotometer 01 2008 ...
Страница 266: ...264 SP600 Spectrophotometer 01 2008 ...
Страница 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Страница 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...
Страница 333: ...331 SP600 Spectrophotometer 01 2008 ...