33
SP600 Spectrophotometer 01/2008
1.1 Methods
Sample preparation: Reaction times must
be exactly kept !
1. Prepare the
dry
reaction apparatus (note 4) and place
it under a fume hood (toxic fumes!).
2. Pipette
50 ml of water sample
into a 100 ml Erlen-
meyer flask (NS 29/32).
3.
30 ml Sulfuric Acid
,
2.0 ml Potassium Iodide solution
and
0.3 ml Tin(II)chloride
are added to the water sample.
4. Close the flask and shake, wait for a period of
15
minutes
.
5. Prepare
2.0 g Zinc
.
6. Fill the absorption tube with exact
5.0 ml of absorption
solution
(see picture
and
; use pipette).
7. After end of the 15 minutes reaction time add the
2 g Zinc to the Erlenmeyer flask and
immediately as-
semble
the apparatus with the prepared absorption
tube (see picture
).
8. The reaction starts
(fume hood!)
.
Wait for
60 minutes
reaction time.
Performing test procedure:
9. Fill a clean
20 mm cell
(note 1) with
deionized water
.
10 Place the cell in the sample chamber making sure that
the positioning is correct.
11. Press
ZERO
key.
12. After zeroing remove the cell from the sample chamber.
Empty the cell and dry completely.
13. Fill the cell with the colored absorption solution. (see
picture
).
14. Place the cell in the sample chamber making sure that
the marks are aligned.
15. Press
TEST
key.
The result is shown in the display in mg/l Arsenic.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Содержание SP600
Страница 1: ...Instruction Manual SP600 Spectrophotometer ...
Страница 2: ......
Страница 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Страница 37: ...35 SP600 Spectrophotometer 01 2008 ...
Страница 73: ...71 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Oxidizing agents interfere as they react like Chlorine ...
Страница 83: ...81 SP600 Spectrophotometer 01 2008 ...
Страница 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 203: ...201 SP600 Spectrophotometer 01 2008 ...
Страница 207: ...205 SP600 Spectrophotometer 01 2008 ...
Страница 231: ...229 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Only ortho Phosphate ions react 2 P PO4 P2 O5 ...
Страница 245: ...243 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will be given ...
Страница 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 251: ...249 SP600 Spectrophotometer 01 2008 ...
Страница 266: ...264 SP600 Spectrophotometer 01 2008 ...
Страница 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Страница 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...
Страница 333: ...331 SP600 Spectrophotometer 01 2008 ...