background image

A41.1950 Operating Microscope, Dual Head, Motor Zoom, 3x~16x. For Neurosurgery, Brain Surgery, ENT 

 

16 

7  Technical specifications

 

(adopt adjustable-pitch observing system ) 

 

Forms of zooming: Variable power optical system with parallel optical path, stepless 

tu rn -d ru m  zoo mi ng

 system   

 

Big focal length of objective:300mm; 

 

Focal length of television interface objective: 75mm, CCD target size: 6mm×4.5mm   

 

Eyepiece magnification: 12.5× 

 

Diaphragm of eyepiece: Φ16 mm   

 

Combination system magnification sees the next table 

 

Illumination lamp: DC, 12V 100W, with variable brightness   

 

Size of light spot: Φ 50mm   

 

Illumination intensity on the surface of object ≥60,000 lx   

 

Non-uniformity of object plane illumination

20% 

 

Z direction focusing scope: ±25mm, Z direction focusing speed:

33s/whole path 

 

Maximum expansion radius: 1100mm 

 

The maximum rotary angle: 360°   

 

Cantilever adjustment angle: -30°~+30 °,lifting scope:0.6m   

 

Weight load on the cantilever: 3kg~12kg   

 

Multifunctional footstep functions: focusing automatically, 

 

Power supply :DC, 220 V ±22V 300W 

 

Dimension: 1350mm×800mm×1600mm     

 

Weight: 115kg   

 

Table 1 

Combination magnifications (focal length of primary objective: 175 mm) 

 

Magnification

 

0.6 

0.8 

1.0 

1.2 

1.4 

1.6 

1.8 

2.0 

2.2 

Total 

magnification 

of 

chief operator microscope

 

7.7 

10.3 

12.8 

15.3 

17.9 

20.5 

23.1 

25.6 

28.1 

Field  of  view  of  chief 
operator microscope

 

26 

19.5 

15.6 

13 

11.1 

9.7 

8.7 

7.8 

7.1 

7.2  Combined system magnification and the conversion of field of view 

For  changing  objective  with  different  focal  length,  the  calculation  relationship  of  total 
magnification of continuous varifocus objective is shown as follows: The total magnification 
of the continuous zooming system has the following relationship:   

  systematic magnifying factor     

Total magnification is as:   

Γ

5magnification

=1993.8*T/f

objective   

Then the field of view is as: 

 

Φ

5field

=0.099* f

objective

/T 

 

8  Directional messages for the customer 

8.1  Routine debugging   

If  fault  occurs  in  your  operation  microscope,  please  inspect  the  instrument  according  to  the 
following  table  and  carry  out  troubleshooting  activities.  If  it  fails  to  remove  troubles,  please 
connect our agent or the Center of After-sale Service 

 

Содержание A41.1950

Страница 1: ...btain satisfactory performance please study this instruction manual thoroughly before start to use your instrument A41 1950 Operating Microscope Dual Head Motor Zoom 3x 16x For Neurosurgery Brain Surg...

Страница 2: ...g the malfunction If you can not solve it by yourselves please contact our agents or the After sale Service Department When the operation finished turn off all of the switches and power off Lid the ob...

Страница 3: ...us damage or operation accident caused by improper operation or application to be carried out do not accord with the requirement of the Operation Manual or professional ability Apparatus damage and me...

Страница 4: ...ce force of the cantilever contrarily will reduce the balance force Tighten the knob will fix the cantilever in the range you need safety slider secure the arm movement within safe distance The power...

Страница 5: ...sal arm 8 4 5 Cantalever 9 4 6 Structure and functions of microscope head 10 4 7 Chief operator observed system 11 4 8 Video recording photography and image processing 13 5 Basic operation 13 5 1 Insp...

Страница 6: ...r chief operator observing system can be choose the one adopts adjustable pitch observing system another adopts 0 observing system The Apparatus has functions of electrical focusing and manual zooming...

Страница 7: ...50 Operating Microscope Dual Head Motor Zoom 3x 16x For Neurosurgery Brain Surgery ENT 2 2 General structures Fig 2 11 12 13 14 1 2 3 4 5 6 7 9 10 8 15 16 17 18 19 20 21 22 23 24 26 27 25 28 29 30 31...

Страница 8: ...1 In The Package The microscope is packed into 4 boxes Box 1 for microscope observing system including microscope head 1 piece chief operator observing system 1piece five spare lamps three lamp frame...

Страница 9: ...t the socket refer to Fig 5 connect the fan plug Connect controled cable conect illumination power line and tighten screw Connect the video plug and push the electronic controlled box into the stand a...

Страница 10: ...nsert it 15 you have to pull the locking pin 19 until you fix the locking nut 21 Fig 6 Step 5 Installation of the Chief operator observing system 16 Align beam splitter to the place pin at the top sur...

Страница 11: ...nob 27 in such a way that the head can rotate in a scope of 30o easily Adjustment of cantalever balance force Loosen the locking knob 24 of cantalever 26 hold the operating handle to move microscope u...

Страница 12: ...s 8 functions to control the electric part of entire instrument and which is convenient for doctor operation There are three switches on the footstep control used for Obliquity control rod 4 focusing...

Страница 13: ...nuous regulation of brightness of cold light source Device 9 is the fuse cover there are some fuses inside Unloading the fuse cover can replace fuse conveniently Additionally the main stand and electr...

Страница 14: ...To moving the slider in arrowhead direction will increase the minimum limited location and contrarily will reduce it After moving the slider please tighten safety valve 25 to fix the slider The instru...

Страница 15: ...Seat of light source 14 Zooming body 15 handle 16 Pitching mechanics 17 Rocker arm 18 Rotor 19 Electronic left right deflection mechanics The head of A41 1950 surgical microscope consists of the pitc...

Страница 16: ...tional objectives such as 200mm 250mm 300mm 350mm and 400mm could be provided in meeting with the need of operation based on customer s request The front of the zooming body 14 has one window 3 indica...

Страница 17: ...rosurgery Brain Surgery ENT 12 Fig 20 variable angle binocular tube 1 Binocular eyepiece 2 Oblique prism case 3 Variable angle Prism case 4 Locked knob 5 Prism case 6 Interface 7 knob for adjusting pu...

Страница 18: ...tific research teaching and long distance medical treatment as shown in Fig 16 The connection of A41 1950 Surgical Microscope with CCD interface and camera and camera will result in continuously zoomi...

Страница 19: ...ation If picture is not clear now adjust eyepiece until the picture become clear Repeat such operation 2 3 times till images at high and low magnifications become distinct Then regulate the other eyep...

Страница 20: ...gas with soda acid If the microscope will not be used for a long time the power socket should be taken down from power supply Seal up microscope head with plastic bag with moisture proof silica gel in...

Страница 21: ...utomatically Power supply DC 220 V 22V 300W Dimension 1350mm 800mm 1600mm Weight 115kg Table 1 Combination magnifications focal length of primary objective 175 mm Magnification 0 6 0 8 1 0 1 2 1 4 1 6...

Страница 22: ...a file make reliable contact Check up whether the fuse of F2 F3 is good Note Fuse F2 and F3 should be changed at the same time Focusing disabled Check up whether the contact of Foot controller socket...

Страница 23: ...g the instrument that exceeds warranty period We have built sale archives Timely reply and processing results will be offered if the customer calls or writes to us We will sincerely study and adopt yo...

Отзывы: