A41.1950 Operating Microscope, Dual Head, Motor Zoom, 3x~16x. For Neurosurgery, Brain Surgery, ENT
7
Fig.9
Step 8:
Correctly connect focusing cable (22),
Obliquity control cable (23), foot controller socket
(8) and the optical fiber (17) (Refer to Section 4.4:.Transversal arm); Insert lamp frame (29)
into the transversal arm (30).
Until this step, the installation of the microscope mainframe is completed.
Step 9:
Both ends of beam splitter can be connected at the same time with television interface and
photo interface. The connection is carried out quickly with treaded ring in line with the notch
after removing the caps at both ends or at one end of the beam splitter.
Notice:
When adding attachment in the head body, the balance force should be raised
properly.
4 Functions of each part
4.1 Multifunctional Foot controller
Multifunctional foot controller has 8 functions to control the electric part of entire instrument and
which is convenient for doctor operation. There are three switches on the footstep control used for
Obliquity control rod (4) , focusing control pedal (2) and
Zooming control pedal(3) respectively.
The control bar (4) is a footstep control device for X-Y
deflection
control in four
directions, it can stop the head in any
Obliquity you want.
Foot stepping after and fore
realizes head deflect in Y direction, stepping left and right for head deflect in X
direction.(As Fig11 shows in 4.6.) No stepping means no head movement.
The
focusing control pedal
(2) is used for focusing chief operator microscope in
longitudinal direction. Up stepping will raise the microscope and down stepping will
drop the microscope. No stepping means stop the head of microscope. It also has
position protection and warning functions in longitudinal direction.
The
Zooming control pedal(3) is used for
adjusting the microscope in any magnification
you want.
27.force balance knob
26.cantalever
24.Locking knob