A41.1950 Operating Microscope, Dual Head, Motor Zoom, 3x~16x. For Neurosurgery, Brain Surgery, ENT
II
Thank you for using our A41.1950 Surgical Microscope
★
Preface
The major content of "The Operation Manual of A41.1950 Surgical Microscope "
consists of the following sections: Product overview, Method of Installing and Simply
operation, Basic functions, Maintenances and troubleshooting, Technical specifications, and
circuit diagram of this apparatus.
★
Safety Guidelines
Please read the manual carefully before using the microscope. Because the safety of
patient, operator and instrument can be guaranteed, Only under the correct installing and
operation. You can get more relevant information from our agents or the After-sale Service
Department.
The surgical microscope is a precise optical apparatus. During operation or storage,
do not expose apparatus under the environment of high temperature, high humidity
and much dust. The instrument can not be exposed in sunlight directly.
When operating the instrument
,
please inspect all the cables whether they are
connected correctly and reliably.
Don’t touch the surface of lens with finger or hard object.
If there are some malfunctions with your apparatus, first please look up the Section
“
8.1
General Troubleshooting Guide” to find some ways to eliminating the
malfunction. If you can not solve it by yourselves, please contact our agents or the
After-sale Service Department.
When the operation finished, turn off all of the switches and power off. Lid the
objective and eyepiece, and cover the dustproof slipcover on it.
This manual contains notices which you should observe to ensure your own personal safety, as
well as to protect the product and connected equipment. These notices are highlighted in the
manual by a warning triangle.
Warning indicates a potentially hazardous situation which. if not avoided, could result
in death or serious injury.
★
Notices for operation
The power socket provided by the customer for the equipment must accord with
international standard (3-hole socket) with earthen protection function, Otherwise
cannot be used.
The biggest balance weight of the cantilever is 13kg (maximal load), overload will
damage the instrument. When balancing force is little, the balance of operation head
will be destroyed. Careless operation will lead to accident. When the balance force is
exceptional, hand-operating feeling is not comfortable. Tuning the knob in counter-clock-wise
direction can increase the balance force of the cantilever and vice versa in clockwise direction
tuning the knob. (Detailed content refers to Section
5.3
Operation of the microscope frames
)
Loose the locking knob to focus the lens and then to tighten the locking knob and fix
the cantilever to prevent the arm from moving during operation. (Detailed content see