A41.1950 Operating Microscope, Dual Head, Motor Zoom, 3x~16x. For Neurosurgery, Brain Surgery, ENT
5
Fig 5
Step 4:
Insert the Main body of the head (15) into the hole of the head joint (18) and fix it
with the locking nut(21). As Fig.6 shows.
Attention : Before you insert it(15), be sure that
you have
unscrew the locking knob(20). And when you insert it(15), you have to pull the locking
pin(19) until you fix the locking nut(21).
Fig.6
Step 5:
Installation of the Chief operator observing system (16). Align beam splitter to the
place-pin at the top surface of zooming body, then turn the obliquity of eyepiece to the biggest
degree (60
°
),and then align eyepiece part onto place-pin on beam splitter, carefully put
eyepiece part onto the interface, finally fix it with locking screw. As the Fig.9 shows.
21.Locking nut
20. Locking knob
15. Main body of the head
19.Locking pin
18.
Head joint