3-8
Section 3
Setting Value Obtaining and Changing Command
ZS Series
CompoWay/F Communication Command Reference
Section 3
Unit No.
and P
a
rameter No.
■
Measurement Setting 1 (Settings per TASK) (Note 2)
Unit No.
(Note 3)
Data No.
Parameter
Setting range/
output range
Remarks
29h 00h
Scaling mode
0: OFF
1: ON
29h 01h
Span value
-20000 to 20000
(X0.0001)
29h 02h
Offset value
-999999999 to 999999999 (nm)
2Ah 02h
SMOOTH
0: OFF
1: ON
2Bh 02h
Average number
of times
0: 1
1: 2
2: 4
3: 8
4: 16
5: 32
6: 64
7: 128
8: 256
9: 512
10: 1024
11: 2048
12: 4096
2Ch 02h
Differentiation mode
0: OFF
1: ON
2Ch 03h
Number
of
differentiation cycles
1 to 5000 (ms)
2Dh 02h
Hold mode
0: OFF
1: PEAK
2: BOTTOM
3: P-P
4: AVERAGE
5: SAMPLE
2Dh 03h
Trigger method
0: External
1: Self-up trigger
2: Self-down trigger
2Dh 04h
Trigger level
-999999999 to 999999999 (nm)
2Dh 05h
Trigger hysteresis
0 to 999999999 (nm)
2Dh 06h
Trigger delay (ms)
0 to 5000 (ms)
2Dh 07h
Sampling period (ms)
1 to 5000 (ms)
2Dh 08h
Trigger delay mode
0: OFF
1: ON
2Eh
05h
Offset value for zero
reset
-999999999 to 999999999 (nm)
2Eh 07h
Zero reset mode
0: REAL
1: HOLD
30h
02h
LOW threshold value
-999999999 to 999999999 (nm)
30h
03h
HIGH threshold value
-999999999 to 999999999 (nm)