3-3
Section 3
Setting Value Obtaining and Changing Command
ZS Series
CompoWay/F Communication Command Reference
Section 3
Unit No.
and P
a
rameter No.
Example3: When obtaining a measurement result of ZS-MDC (0CH) TASK3:
02h 01h (C000h + 20h (Data No.)) 58h (Unit No.) 00h (CH No.) (8000h + 1h (Number of elements))
In the command section, the value is shown as below:
02h 01h C020h 5800h 8001h
To this command, if the response is shown as below:
02h 01h C020h 5800h 8001h FFF0BDC0h
It indicates as -1000000nm (= -1mm) because FFF0BDC0h = -1000000.
(A negative value is expressed as a complement of two.)