38
Run-Mode Measurement Process
Setting about Individual Data (Serial/SD-Card) [ASCII Format]
MENU mode - [System] - [Output] - [Date format (Serial)/(SD Card)]
Output format
When “Display at time” is ON, the time information is prefixed.
Setup Item
Description
Output form
Selects ASCII.
Digits of integer
Sets the number of output digits of the integer section.
When "0" is selected, all of the digits of the data are output
shifted to the left.
When there are fewer data digits than the preset number of digits,
"0" is inserted in free digits.
When there are more data digits than the preset number of digits,
"9" is output for the preset number of digits.
Range: 0 to 8 (default value: 8)
Digits of decimal
Sets the number of output digits past the decimal point.
When "0" is selected, digits past the decimal point are rounded
up to the nearest integer.
Range: 0 to 3 (default value: 3)
Field separator
Sets the separator between individual output data items.
Range: None, Comma (default value), Tab, Space, Semicolon
Decimal
separator
Sets the number of digits past the decimal point.
Range: None, Period (default value), Comma
Record separator Sets the separator between individual output data records.
Range: None, Comma, Tab, Space, Delimiter (default value)
Display at time
Prefixes the output data with time information.
Range: OFF (default value), ON
File name
Sets the name of the output file. (only when SD card is selected)
The directory "OUTFILE" is automatically created in the root
directory of the SD card. Output files are stored in this directory.
< Measurement value of data 0 >
,
Field separator
Record separator
< Measurement value of data 1>... < Measurement value of data 31 > Delimiter
<Year/month/day>
,
<Hours/minutes/seconds>
,
<Measurement value of data 0>
,
<Measurement value of data 1> ... <Measurement value of data 31> Delimiter
Example:
On August 1st
,
2007
→
2007/8/1
Example:
At 1 minute
,
20 seconds past 12 am
→
12:01:20
Z01E-EN-01+ZFX+SettingGuide.book Seite 38 Mittwoch, 13. Februar 2008 2:38 14