21.Error List
E81
Measurement
Value
Aberration.
The reading was not
accurately taken.
The result of the
pulse wave
measurement will
not be displayed.
Detect an abnormal
measurement value.
• There is a possibility that the AI
is less than 30% or larger than
150%. Confirm the measured
pulse waveform. After reviewing
Section 9.2. and Section 14.,
retake the measurement.
E82
Abnormal
average value
calculation.
An irregular pulse wave
or similar problem
prevented the average
value from being
measured.
• There is a possibility that the AI
pulse wave sensor is not on the
artery. After reviewing Section
9.2. and Section 14., retake the
measurement.
E85
Position Error.
The AI pulse wave
sensor is not
positioned on the
artery.
Slide the AI pulse
wave sensor unit in
direction ‘A’ and
restart
measurement.
An artery is not under
the sensor.
• Slide the AI pulse wave sensor
unit one or two scales in the
indicated direction and restart
the measurement. For complete
information refer to Section
12.1.
E86
Position Error.
The AI pulse wave
sensor is not
positioned on the
artery.
Slide the AI pulse
wave sensor unit in
direction ‘B’ and
restart
measurement.
An artery is not under
the sensor.
• Slide the AI pulse wave sensor
unit one or two scales in the
indicated direction and restart
the measurement. For complete
information refer to Section
12.1.
E87
Unstable Blood
Pressure
Measurement -
Reliability
Compromised
Unstable blood
pressure
measurement.
Measurement results
may be
compromised.
The pulse wave sensor
detected a blood
pressure fluctuation.
• Ensure that the patient is at rest
and retake the measurement.
E88
Unstable Blood
Pressure
Measurement -
Invalid
Measurement
Unstable blood
pressure
measurement.
Measurement results
may be
compromised.
cSBP cannot be
calculated.
The pulse wave sensor
detected a blood
pressure fluctuation.
• Ensure that the patient is at rest
and retake the measurement.
E89
Body
Movement -
Invalid
Measurement
Unstable blood
pressure
measurement due to
body or device
movement,
Measurement results
may be
compromised.
cSBP cannot be
calculated.
Body movement is
detected.
• Ensure that the patient is at rest
and retake the measurement.
Error
No.
Error Name
Error Message on
Display
Error Content
Action