19.Calibration Measurement
EN
19.Calibration Measurement
It is recommended that you have the monitor inspected every two years to ensure it is accurate
and functioning correctly.
19.1. How to Execute a Test Measurement
You will need:
(1) A mercury blood pressure monitor that is accurately adjusted (including a rubber bulb)*.
(2) 2 T-tubes
(3) 2 air tubes
(4) Something to wrap a cuff around such as a sturdy tube (strong material that is not damaged or
deformed by inflation)
* Even a mercury blood pressure monitor might cause an error in measurement due to a lack of
mercury in the mercury tank, dirt on the glass tube or clogging of the glass tube cap.
1.
Connect the mercury blood pressure monitor, the
rubber bulb, the main unit and the T-tube as shown
in [Fig. 1], to the right
2.
Press the F5 key to enter the [SETUP] screen.
[Fig. 2]
3.
Move the [
] cursor to [METHOD] under [BP
MEASUREMENT] in the [SETTING LIST] column.
[Fig. 3]
[Fig. 1]
T-tube
[Fig. 2]
[Fig. 3]