18.Configuration Options
EN
7.
To change the [DISPLAY TONOGRAM] under [AI
MEASUREMENT]:
1. Move the [
] cursor to [DISPLAY TONOGRAM].
[Fig. 10]
2. Press the SET or [
] button to move the [
] cursor to
[SETTING VALUE].
3. Move the [
] cursor to the desired setting, selecting
from [PATIENT VIEWPOINT] or [OPERATOR
VIEWPOINT]. [Fig. 11]
The default setting is [OPERATOR VIEWPOINT].
4. Press the SET button to accept the setting value.
5. Press the F1 key [EXIT] to return to the default screen.
8.
To change the [MEASUREMENT SEQUENCE] under
[AI AND BP MEASUREMENT]:
1. Move the [
] cursor to [MEASUREMENT SEQUENCE].
[Fig. 12]
2. Press the SET or [
] button to move the [
] cursor to
[SETTING VALUE].
3. Move the [
] cursor to the desired setting, selecting
from [AI
→
BP] or [AI + BP] under [SETTING VALUE].
[Fig. 13]
The default setting is [AI + BP].
4. Press the SET button to accept the set value.
5. Press the F1 key [EXIT] to return to the default screen.
[Fig. 10]
[Fig. 11]
[Fig. 12]
[Fig. 13]