5
Introduction
(continued)
Step One
SPECIFICATIONS
Electrical:
Supply Voltage:
14-28 Vdc
Signal:
4 to 20 mA current loop; 22 mA
max.
Mechanical:
Display Module:
Removable via short cable and
plug
LCD:
6 characters, 10 mm (0.394”)
height
Buttons:
Three, UP / OK / DN
Pin Connection: Display can be removed such
that the open connection allows
the sensor to be LVCN414-SW
software capable.
Enclosure:
Polypropylene head with skirt to
support transducer w/ conduit
Conduit Connection: Single, 1/2 inch NPT
Transducer:
Black, PVDF with FKM gasket
Process Mount:
LVU706: 2" NPT or 2” G
LVU712: 3” NPT or 3” G
Gasket
FKM (G thread only)
Electrical Connections:
Power & Return: 2-pin removable terminal block
Display Connect.: 4-pin removable terminal pins
Non-Display
version.:
Block optional with FOB
* Earth ground on package
external
Environmental:
Process Temp.:
-40 to 80 °C (-40 to 176 °F)
Temp. comp.:
Automatic
Ambient Temp.:
-35 to 60 °C (-31 to 140 °F)
Pressure:
30 psi, derated @ 1.667 psi per
degree C above 25 °C
Enclosure:
NEMA 6P, IP68
Functional:
Range:
LVU706: 6.0m (19.69 feet)
LVU712: 12.0m (39.4 feet)
Dead band:
LVU706: 20cm (8 inches)
LVU712: 45.7cm (18 inches)
Accuracy:
0.2% of Range in air at 20 °C
Resolution:
LVU706: 2mm (0.079 inches)
LVU712: 5mm (0..196 inches)
Beam width:
LVU706: 7.6cm (3 inches)
LVU712: 15.2cm (6 inches)
Menu:
Self-scrolling
Содержание LVU-700
Страница 1: ...1...
Страница 2: ...2...
Страница 6: ...6 Introduction continued Step One DIMENSIONS Dimensions for LVU706 Series Dimensions for LVU712 Series...
Страница 51: ...51...