
8-35
8
Scaling to
Display
Engineering
Units
STRAIN
0
Go back into the set up mode and follow the flow chart below
entering the values in the submenus shown.
MENU
SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF
Skip past
L2CNF
Skip past
L3CNF
Skip past
L4CNF
Skip past
INPUT
Skip past
RDG.CNF
Skip past
RDG SC
Skip past
(Leave as 1.00000)
RDG OF
Skip past
(Leave as 000000)
IN CNF
INP.6=1
Activates IN.SC.OF
(2-point scaling)
IN.SC.OF
INPUT 1
019590
This is where you
READ 1
000000
enter the data
INPUT2
080150
obtained during the
READ 2
004000
calibration run (see above).
DEC PT
FFFFF.F
Decimal point selection
CNT BY
Press the ‘RESET’ button
two times to return to
the run mode
The unit is now completely scaled. If during verifying the
calibration, the display is not correct, you can use the offset
function (“RDG OF” if all values read high or low) or perform
this calibration again.
Содержание DP461-S
Страница 96: ......