9-27
Meter Function Menus
9
9.2.19 SERCNF (Serial Communication Configuration)
Serial communication configuration is used to select:
• no parity, odd parity, or even parity for communications [SER.1]
• 1 stop bit or 2 stop bits [SER.2]
• No parity, 1 stop is not allowed
MENU
MIN/MAX/MENU *
BUTTON
BUTTON
MAIN MENU
SUB MENU
DESCRIPTION
“SERCNF”:
SERIAL COMMUNICATION
CONFIGURATION
Select parity for communications via
RS-232 or RS-485:
“SER.1=0”
No parity
“SER.1=1”
Odd parity
“SER.1=2”
Even parity
Select stop bits for communications via
RS-232 or RS-485:
“SER.2=0”
1 Stop Bit
“SER.2=1”
2 Stop Bits
* The
‘MIN’
button allows you to toggle between SER.1 and SER.2.
The
‘MAX’
button allows you to select the “0”, “1”, or “2” state for each
“SER” condition.
The
‘MENU’
button stores the selected values for each “SER” condition
changed and advances the meter to the next configuration (“ADDRES”).
Every underlined “0”, “1” or “2” is the factory preset value.
NOTE:
Only 7 data bit communication is supported